International Journal of Obstetric Anesthesia
Volume 15, Issue 4 , Pages 290-293, October 2006

Effect of blood pressure instrument and cuff side on blood pressure reading in pregnant women in the lateral recumbent position

  • S.M. Kinsella

      Affiliations

    • Corresponding Author InformationCorrespondence to: Dr. Stephen Michael Kinsella, Department of Anaesthesia, St. Michael’s Hospital, Bristol BS2 8EG, UK. Tel.: +0117 928 5203; fax: +0117 928 5209.

Department of Anaesthesia, St. Michael’s Hospital, Bristol, UK

Accepted 1 March 2006.

Background

Hydrostatic forces affect non-invasive blood pressure measurement in the lateral position. This study assessed the extent of this effect with the mercury column sphygmomanometer and Dinamap oscillometric instrument as well as different recommendations for comparing supine and lateral blood pressure measurements.

Method

Thirty-two term pregnant women were studied in the antenatal clinic. Blood pressure was recorded from both arms in the right lateral and supine recumbent positions, using the sphygmomanometer and Dinamap.

Results

Blood pressure in the uppermost arm while lateral was lower than supine by a mean 10mmHg or more. Systolic, mean and diastolic pressures in the dependent arm while lateral were higher than supine by a mean (SD) 3.1 (6.8)mmHg, 5.6 (6.8)mmHg, and 6.9 (8.7)mmHg using the sphygmomanometer and 3.8 (8.1)mmHg, 3.2 (7.1)mmHg, and 1.9 (5.3)mmHg using the Dinamap. Systolic, mean and diastolic pressure values calculated as the average taken from both arms in the lateral position were lower than supine by a mean (SD) 3.5 (7.5)mmHg, 3.9 (4.7) mmHg, and 4.1 (5.8)mmHg using the sphygmomanometer and 4.6 (6.0)mmHg, 4.9 (4.4)mmHg, and 4.8 (4.4)mmHg using the Dinamap. Corresponding blood pressure readings were always higher using the Dinamap than the sphygmomanometer.

Conclusions

In normotensive non-labouring term pregnant women, the use of the dependent arm or an average blood pressure from both arms while in the lateral position will give a closer reading to supine blood pressure than the use of the uppermost arm. However, use of the dependent arm is simpler.

Keywords: Blood pressure, Pregnancy, Position, Supine, Lateral, Cuff position, Instrument, Measurement

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0959-289X(06)00097-5

doi:10.1016/j.ijoa.2006.03.010

International Journal of Obstetric Anesthesia
Volume 15, Issue 4 , Pages 290-293, October 2006