A prospective observational study of the change in regional cerebral oxygen saturation during cesarean delivery in women receiving phenylephrine prophylaxis for spinal hypotension

Published:September 20, 2018DOI:https://doi.org/10.1016/j.ijoa.2018.09.005

      Highlights

      • Near-infrared spectroscopy (NIRS) measures regional cerebral oxygen saturation.
      • Prophylactic phenylephrine for cesarean section has cerebral haemodynamic effects.
      • Spinal anesthesia with phenylephrine infusion correlated with decreased NIRS values.
      • Maximum decrease in NIRS value was five minutes after initiation of spinal block.
      • The clinical significance of these changes remains to be established.

      Abstract

      Background

      Spinal hypotension causes decreased regional cerebral oxygen saturation (ScO2) in women undergoing cesarean delivery. In this study we aimed to measure the change in ScO2 using near infrared spectroscopy in women receiving a prophylactic phenylephrine infusion during cesarean delivery under spinal anesthesia.

      Methods

      This was a prospective, observational cohort study. Fifty-three women had ScO2 measurements at the following time points: preoperatively, in the supine position with 30° of left lateral tilt; one and five minutes after spinal anesthesia; at the time of skin incision; immediately after delivery; one minute after commencing the oxytocin infusion; at completion of surgery, and one hour after surgery. Spinal anesthesia and a prophylactic phenylephrine infusion were administered according to a standard treatment protocol. Statistical analysis used the Wilcoxon Signed Rank test with Bonferroni’s correction for multiple comparisons.

      Results

      Blood pressure was maintained within 20% of baseline throughout surgery. The baseline mean (range) ScO2 was 61.5% (54.0–66.3%). It decreased significantly at all subsequent measurement points. The maximum decrease was five minutes after spinal anesthesia. Thirty-four (64.2%) of the parturients exhibited ScO2 values <20% of baseline, or a decrease to below an absolute value of 50%. There was no significant correlation between systolic blood pressure and mean ScO2.

      Conclusion

      Spinal anesthesia with phenylephrine infusion during cesarean delivery is associated with a significant decrease in ScO2 levels, maximal five minutes later. Further studies are required to establish the clinical significance of this finding.

      Keywords

      Introduction

      The incidence of spinal hypotension during cesarean delivery (CD) is as high as 70–80%,
      • Mercier F.J.
      • Auge M.
      • Hoffmann C.
      • Fischer C.
      • Le Gouez A.
      Maternal hypotension during spinal anesthesia for caesarean delivery.
      and may cause dizziness, faintness and nausea and vomiting, secondary to cerebral hypoperfusion.
      • Habib A.S.
      A review of the impact of phenylephrine administration on maternal hemodynamics and maternal and neonatal outcomes in women undergoing cesarean delivery under spinal anesthesia.
      Near infrared spectroscopy (NIRS) is a direct non-invasive spectroscopic method that uses the near infrared portion of the electromagnetic spectrum (800–2500 nm) to measure blood oxygenation in the frontal lobe microvasculature.
      • Calderon-Arnulphi M.
      • Alaraj A.
      • Slavin K.V.
      Near infrared technology in neuroscience: past, present and future.
      Its advantages include simple bedside performance and lack of interference with other measurements.
      • Calderon-Arnulphi M.
      • Alaraj A.
      • Slavin K.V.
      Near infrared technology in neuroscience: past, present and future.
      Near infrared spectroscopy has previously demonstrated that spinal hypotension causes decreased brain oxygen saturation during CD.
      • Sun S.
      • Liu N.H.
      • Huang S.Q.
      Role of cerebral oxygenation for prediction of hypotension after spinal anesthesia for caesarean section.
      • Hirose N.
      • Kondo Y.
      • Maeda T.
      • Suzuki T.
      • Yoshino A.
      Relationship between regional cerebral blood volume and oxygenation and blood pressure during spinal anesthesia in women undergoing cesarean section.
      When ephedrine was administered to treat spinal hypotension, cerebral oxygen saturation (ScO2) returned to baseline values.
      • Berlac P.A.
      • Rasmussen Y.H.
      Per-operative cerebral near-infrared spectroscopy (NIRS) predicts maternal hypotension during elective caesarean delivery in spinal anaesthesia.
      • Foss V.T.
      • Christensen R.
      • Rokamp K.Z.
      • Nissen P.
      • Secher N.H.
      • Nielsen
      HB. Effect of phenylephrine vs. ephedrine on frontal lobe oxygenation during caesarean section with spinal anesthesia: an open label randomized controlled trial.
      Current standard practice is to use a prophylactic phenylephrine infusion to prevent spinal hypotension during CD.
      • Loubert C.
      Fluid and vasopressor management for Cesarean delivery under spinal anesthesia: continuing professional development.
      • Kinsella S.M.
      • Carvalho B.
      • Dyer R.A.
      • et al.
      International consensus statement on the management of hypotension with vasopressors during caesarean section under spinal anaesthesia.
      There are few studies discussing the effect of phenylephrine on ScO2.
      • Foss V.T.
      • Christensen R.
      • Rokamp K.Z.
      • Nissen P.
      • Secher N.H.
      • Nielsen
      HB. Effect of phenylephrine vs. ephedrine on frontal lobe oxygenation during caesarean section with spinal anesthesia: an open label randomized controlled trial.
      • Meng L.
      • Cannesson M.
      • Alexander B.S.
      • et al.
      Effect of phenylephrine and ephedrine bolus treatment on cerebral oxygenation in anaesthetized patients.
      • Nissen P.
      • Brassard P.
      • Jørgensen T.B.
      • Secher N.H.
      Phenylephrine but not ephedrine reduces frontal lobe oxygenation following anesthesia-induced hypotension.
      In a randomized controlled study that evaluated the fall in average frontal lobe ScO2 over 60 minutes after administration of spinal anesthesia, a prophylactic phenylephrine infusion and a prophylactic ephedrine infusion were compared in 24 women. Phenylephrine infusion was associated with a decrease in mean ScO2 of −8.6 ± 2.8%,
      • Clemmesen C.G.
      • Pedersen L.M.
      • Hougaard S.
      • et al.
      Cerebral oximetry during preoperative resuscitation in elderly patients with hip fracture: a prospective observational study.
      despite similar blood pressure maintenance in both groups, while mean ScO2 using ephedrine infusion was unchanged.
      We performed a prospective observational pilot study to investigate the change in mean frontal lobe ScO2 at different time points intra- and postoperatively in women receiving a prophylactic phenylephrine infusion during CD. Based on results from a preliminary study in our hospital, the primary aim of the study was to measure mean ScO2 levels five minutes after initiation of spinal anesthesia. The secondary aims were to measure: (1) change in mean ScO2 from baseline at the following time points: 1-min after initiation of spinal anesthesia (1-min), 5-min after spinal anesthesia (5-min), at skin incision (SI), immediately after baby delivery (BD), 1-min after initiation of oxytocin infusion (OX), at surgery completion (SE) and one hour after surgery in the post-anesthesia care unit (PACU1); (2) the decrease of mean frontal ScO2 below the theoretical cerebral hypoperfusion threshold; (3) systolic blood pressure (SBP) during surgery compared to baseline; (4) the correlation between SBP and mean frontal lobe ScO2.

      Methods

      This prospective observational study was performed at the Beilinson Campus of the Rabin Medical Centre, Petah Tikva, Israel, a tertiary university hospital with 10 000 deliveries per annum, between October 2015 and March 2016. The Institutional Review Board approved this study and all participants provided written informed consent. As per institutional policy, the study was registered on clinical trials.gov (registration number NCT02473978).
      Consecutive women undergoing elective CD under spinal anesthesia with a prophylactic phenylephrine infusion were included. Exclusion criteria were women under the age of 18 years, American Society of Anesthesiologists (ASA) physical score >2, preeclampsia or chronic hypertension, unfamiliarity with the Hebrew language, and allergy to phenylephrine. Women in whom the surgeon assessed significant intraoperative hemorrhage, or who required blood products or methergine for uterine atony, were excluded from analysis.
      In the ward, an intravenous cannula (18-gauge) was inserted 2.5 h before surgery and 500 mL of Ringer’s Lactate was infused over two hours. Sodium citrate (30 mL) and Ringer’s Lactate solution (500 mL) were given 30 minutes before surgery.
      Women were placed supine with left uterine displacement, using a wedge, in the operating room (OR). Standard monitoring with non-invasive blood pressure (NIBP), pulse oximetry and electrocardiogram (S/5 Anesthesia Delivery Unit, Instrumentarium Corp., Datex-Ohmeda, Finland) was instituted. Baseline NIBP was measured in the 30° left lateral supine position on arrival in the OR and at 1-min intervals until the end of surgery. Near infrared spectroscopy sensors (INVOS Cerebral/Somatic Oximeter|Covidien, Boulder Colorado USA) were connected to the right and left side of the forehead, immediately below the hairline, according to the manufacturer’s guidelines, to measure right- and left-sided (ScO2R; ScO2L) frontal cerebral oxygen saturation. No women received routine supplemental oxygen.
      Spinal anesthesia was performed in the sitting position. A 26-gauge needle (B Braun Pencil Point Spinal Needle) was inserted at the L3–L4 or L4–L5 interspace, and hyperbaric bupivacaine 12 mg (AstraZeneca, England), fentanyl 20 µg (Panpharma, France) and preservative-free morphine sulphate 0.1 mg (Rafa Laboratories, Israel) were injected. After injection, women were placed in the supine position with left uterine displacement.
      Immediately after completion of spinal injection, a phenylephrine infusion (Omega, Canada) was administered at an initial rate of 50 µg/min. The phenylephrine dose was titrated according to a standard protocol: if SBP fell 20% from baseline or below 90 mmHg, or faintness, dizziness, nausea or vomiting occurred, the initial dose was increased by 20%; if SBP increased more than 20% from baseline the initial dose was decreased by 20%. The phenylephrine infusion was continued throughout surgery and was not tapered after delivery, due to our institutional protocol of uterine exteriorization and associated nausea and vomiting seen at this time. The phenylephrine infusion was stopped after skin stapling. The sensory level to pinprick test was checked and surgery commenced when the block had reached the T4 dermatome. Immediately after delivery, oxytocin 20 IU in 1L Ringer’s Lactate was infused at 200 mL/h. No other uterotonics were given. All parturients received dexamethasone 4 mg (Omega, Canada) and ondansetron 4 mg (Fresenius Kabi, Germany) after delivery.
      In the PACU, the Ringer’s Lactate infusion with oxytocin was changed to a bag containing 10 IU/L and infused at 200 mL/h. All women received intravenous tramadol 100 mg (Grunenthal GMBH, Germany) with 10 mg metoclopramide (Rafa laboratories, Israel), 1000 mg intravenous paracetamol (Bristol Myers Squibb, USA) and intramuscular diclofenac 75 mg (Teva Pharmaceutical, Israel).
      Demographic data (maternal age, weight, parity, gravidity) and neonatal Apgar score were recorded. Frontal lobe ScO2 was recorded preoperatively in the 30° left lateral supine position on arrival in the OR (BL), 1-min after initiation of spinal anesthesia (1-min), 5-min after spinal anesthesia (5-min), at skin incision (SI), immediately after baby delivery (BD), 1-min after initiation of oxytocin infusion (OX), at surgery completion (SE) and 1 h after surgery in the PACU (PACU1). There was no difference between left- and right-sided cerebral oxygen saturation, so the average of both frontal lobe ScO2 measures was used and recorded as mean cerebral saturation. The theoretical cerebral hypoperfusion threshold was defined as an absolute decrease of frontal lobe ScO2 to <50% or a decrease of more than 20% from baseline on either right or left side of the brain.
      • Casati A.
      • Fanelli G.
      • Pietropaoli P.
      • et al.
      Continuous monitoring of cerebral oxygen saturation in elderly patients undergoing major abdominal surgery minimizes brain exposure to potential hypoxia.
      • Murkin J.M.
      • Adams S.J.
      • Novick R.J.
      • et al.
      Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study.
      The NIBP values at these time points were also recorded. Total phenylephrine dose was noted.
      The primary outcome of this pilot study was the change from baseline of mean frontal lobe ScO2, as measured by NIRS at 5-min. Secondary outcomes were:
      • 1.
        Change from baseline of mean frontal lobe ScO2 at all other time points.
      • 2.
        Decrease of frontal lobe ScO2 below the cerebral hypoperfusion threshold.
      • 3.
        Systolic blood pressure measurements throughout surgery compared to baseline.
      • 4.
        Correlation between systolic blood pressure and mean frontal lobe ScO2.
      Based on previous research to detect a decrease in mean frontal lobe ScO2 readings from baseline to lowest mean ScO2 of more than 6%, assuming a common standard deviation of 8% and allowing for 20% of potential dropouts, 56 women were required to obtain a power of a least 0.8 (β = 20%; two-tailed α = 5%).
      • Calderon-Arnulphi M.
      • Alaraj A.
      • Slavin K.V.
      Near infrared technology in neuroscience: past, present and future.
      • Sun S.
      • Liu N.H.
      • Huang S.Q.
      Role of cerebral oxygenation for prediction of hypotension after spinal anesthesia for caesarean section.
      Statistical analysis was performed using the SPSS software version 23.0 (IBM, Chicago, IL). Variables were analyzed with statistical tests according to the Kolmogorov–Smirnov test of normality. According to the Kolmogorov–Smirnov test of normality, both normally and non-normally distributed data were observed (Supplementary file). Nonparametric tests were therefore used for all data.
      The primary outcome; comparison of mean baseline ScO2 with other time points; and the decrease of ScO2 below the theoretical cerebral hypoperfusion threshold, were measured using the Wilcoxon Signed Rank test with Bonferroni’s correction for multiple comparisons and presented as boxplots.
      The SBP measurements were compared with baseline using the Wilcoxon Signed Rank test with Bonferroni’s correction for multiple comparisons; and data are presented as boxplots. The correlation between SBP and mean frontal lobe ScO2 was measured using the Spearman correlation test. Seven comparisons were used for each variable: baseline to 1-min; baseline to 5-min; baseline to SI; baseline to BD; baseline to OX; baseline to SE; baseline to PACU1. The P-value was corrected as per 0.05/7 = 0.007.

      Results

      We enrolled 59 women, and three were excluded (Fig. 1). The outcomes for 56 women are reported. Demographic data, neonatal data, and total phenylephrine dosage are shown in Table 1.
      Figure thumbnail gr1
      Fig. 1Study patient flowchart. NIRS: near infrared spectroscopy
      Table 1Maternal and neonatal demographic data, duration of surgery, phenylephrine dose and length of stay in the postoperative care unit
      Parametern=53
      Age (y)35 (29.5–38)
      Weight (kg)74 (66–85)
      Parity (n)1 (0.5–2)
      Gravity (n)3 (2–4)
      Gestation week38.1 (37.4–38.5)
      PACU (min)173 (151–228)
      Surgery (min)24 (19.5–29.5)
      Phenylephrine dose (mg)1.5 (1.28–1.78)
      Umbilical artery pH7.37 (7.35–7.39)
      Apgar 1-min9 (9–9)
      Apgar 5-min10 (10–10)
      Data presented as median (25–75th percentiles). PACU: post-anaesthesia care unit.
      Oxygen saturation was 96–98% during the measurement period.
      For the primary outcome, at 5-min after spinal anesthesia, the mean frontal lobe ScO2 had decreased significantly compared to baseline, from 61.5% (54–66.3) to 55% (48–60.8) (P <0.005).
      Secondary outcomes are listed in Fig. 2. Compared to baseline, mean frontal lobe ScO2 decreased significantly at all time points during surgery, starting five minutes after the spinal. At the end of surgery, after discontinuation of the phenylephrine infusion, there was a non-significant increase in mean frontal lobe ScO2 compared to baseline, however in the PACU one hour after completion of surgery, mean frontal lobe ScO2 was again significantly reduced compared to baseline.
      Figure thumbnail gr2
      Fig. 2Cerebral oxygen saturation according to time. BL: baseline; SI: skin incision; BD: baby delivery; OX: oxytocin infusion; SE: surgery completion; PACU: postoperative anesthesia care unit
      For the theoretical cerebral hypoperfusion threshold, 34 (64.2%) women had cerebral hypoperfusion at least once during the study period (Table 2). Twenty-six parturients (49.1%) had cerebral hypoperfusion five minutes after spinal anesthesia. In the PACU, one hour after surgery, seventeen (32.1%) women remained below the theoretical cerebral hypoperfusion threshold.
      Table 2Cerebral hypoperfusion threshold at different measurement points
      Measurement pointCerebral hypoperfusion threshold n=53
      Any sideLeft-sidedRight-sided
      1-min4 (7.6)4 (7.6)2 (3.8)
      5-min26 (49.1)21 (39.6)19 (35.9)
      SI24 (45.3)20 (37.7)16 (30.2)
      BD16 (30.2)12 (22.6)13 (24.5)
      OX13 (24.5)11 (20.8)11 (20.8)
      SE8 (15.1)5 (9.4)6 (11.3)
      PACU117 (32.1)14 (26.4)11 (20.8)
      Total number of patients34 (64.2)29 (54.7)27 (50.9)
      Cerebral hypoperfusion threshold = absolute decrease of cerebral saturation <50% or decrease more than 20% from baseline. Data presented as number (percent). Points of measurement:
      1-min – one minute after initiation of spinal anesthesia. 5-min – five minutes after spinal anesthesia; SI: skin incision; BD: baby delivered; OX: one minute after beginning of oxytocin infusion; SE: surgery completion; PACU1: one hour after surgery.
      Systolic blood pressure measurements are shown in Fig. 3. Baseline SBP was 121 (114.5–137) mmHg, and no statistically significant decrease in blood pressure occurred throughout surgery. However, at the end of surgery, after ScO2, was discontinued phenylephrine infusion was discontinued, and SBP decreased significantly (P <0.007) from 121 (114.5–137) to 106 (92–114.5) mmHg. One hour after surgery, blood pressure remained significantly lower (P <0.007) than the baseline (from 121 (114.5–137) to 116 (109.5–124) mmHg).
      Figure thumbnail gr3
      Fig. 3Systolic blood pressure measurements. BL: baseline; SI: skin incision; BD: baby delivery; OX: oxytocin infusion; SE: surgery completion; PACU: postoperative anesthesia care unit
      There was no significant correlation between SBP and mean frontal lobe ScO2 (RS 0.03 for SBP, P=0.6), although our study was not powered to evaluate this.

      Discussion

      In this prospective pilot study, NIRS was used to measure mean frontal lobe ScO2 in women receiving a prophylactic phenylephrine infusion. Despite preservation of blood pressure to within 20% of baseline measurements, the mean frontal lobe ScO2 significantly decreased after spinal anesthesia and remained below baseline at all measured time-points until the phenylephrine infusion was discontinued. It decreased further one hour postpartum. Mean frontal lobe ScO2 was below the theoretical cerebral hypoperfusion level at some point during surgery in more than a third of women.
      The fall in mean frontal lobe ScO2 corroborates the findings of Foss et al., who showed decreased frontal lobe ScO2 after spinal anesthesia in conjunction with phenylephrine infusion over the 60 minutes after spinal anesthesia.
      • Foss V.T.
      • Christensen R.
      • Rokamp K.Z.
      • Nissen P.
      • Secher N.H.
      • Nielsen
      HB. Effect of phenylephrine vs. ephedrine on frontal lobe oxygenation during caesarean section with spinal anesthesia: an open label randomized controlled trial.
      Our study investigated specific perioperative time-points and the magnitude of the mean frontal lobe ScO2 change at each time-point.
      The mechanism for the decrease in mean frontal lobe ScO2 is unclear and likely multifactorial. In one study measuring ScO2, a third NIRS sensor was placed on the thigh to determine the effect of spinal block on the lower extremities.
      • Fassoulaki A.
      • Paraskeva A.
      • Tsaroucha A.
      Cesarean delivery under spinal anesthesia is associated with decreases in cerebral oxygen saturation as assessed by NIRS: an observational study.
      Spinal anesthesia caused redistribution of blood flow to the lower extremities, leading to decreased mean frontal lobe ScO2, since reduced mean frontal lobe ScO2 values corresponded with increased regional saturation of the thigh.
      A further explanation of the findings may be occult spinal hypotension. Previous studies have shown the relationship of spinal-induced hypotension and decreased ScO2.
      • Foss V.T.
      • Christensen R.
      • Rokamp K.Z.
      • Nissen P.
      • Secher N.H.
      • Nielsen
      HB. Effect of phenylephrine vs. ephedrine on frontal lobe oxygenation during caesarean section with spinal anesthesia: an open label randomized controlled trial.
      • Fassoulaki A.
      • Paraskeva A.
      • Tsaroucha A.
      Cesarean delivery under spinal anesthesia is associated with decreases in cerebral oxygen saturation as assessed by NIRS: an observational study.
      In one study of 35 parturients using low-dose spinal anesthesia, women who developed hypotension had a preceding decrease in mean frontal lobe ScO2.
      • Berlac P.A.
      • Rasmussen Y.H.
      Per-operative cerebral near-infrared spectroscopy (NIRS) predicts maternal hypotension during elective caesarean delivery in spinal anaesthesia.
      In two other studies, women who received spinal anesthesia for CD had decreased mean frontal lobe ScO2 that was inversely correlated with NIBP readings.
      • Fassoulaki A.
      • Paraskeva A.
      • Tsaroucha A.
      Cesarean delivery under spinal anesthesia is associated with decreases in cerebral oxygen saturation as assessed by NIRS: an observational study.
      Although we attempted to maintain blood pressure within a narrow range, it is possible that the blood pressure measured every minute missed intermittent hypotensive episodes. A study comparing intermittent versus continuous NIBP monitoring during CD found that intermittent monitoring every three minutes missed hypotensive episodes in 30% of episodes.
      • Ilies C.
      • Kiskalt H.
      • Siedenhans D.
      • et al.
      Detection of hypotension during Caesarean section with continuous non-invasive arterial pressure device or intermittent oscillometric arterial pressure measurement.
      Another consideration is that intraoperative shivering or movement may have interfered with the reading. Cerebral blood flow changes produced by deafferentation may also be a cause.
      • Sadato N.
      • Zeffiro T.A.
      • Campbell G.
      • Konishi J.
      • Shibasaki H.
      • Hallett M.
      Regional cerebral blood flow changes in motor cortical areas after transient anesthesia of the forearm.
      In early research on the effect of deafferentation on cortical areas activated by movement of proximal muscles, regional cerebral blood flow was measured with positron emission tomography and 15O-labeled water. The authors found that movement-related change in cerebral blood flow was reduced after deafferentation.
      • Sadato N.
      • Zeffiro T.A.
      • Campbell G.
      • Konishi J.
      • Shibasaki H.
      • Hallett M.
      Regional cerebral blood flow changes in motor cortical areas after transient anesthesia of the forearm.
      A final explanation for the decreased mean frontal lobe ScO2 may be the direct effect of phenylephrine. Studies in healthy volunteers receiving phenylephrine infusion showed a decrease in mean frontal lobe ScO2 when compared to the same volunteers who received a saline infusion.
      • Ogoh S.
      • Sato K.
      • Fisher J.P.
      • Seifert T.
      • Overgaard M.
      • Secher N.H.
      The effect of phenylephrine on arterial and venous cerebral blood flow in healthy subjects.
      Anesthetized, non-obstetric patients who received phenylephrine boluses to correct hypotension demonstrated a paradoxical decrease in frontal lobe ScO2, apparently from a direct effect.
      • Meng L.
      • Cannesson M.
      • Alexander B.S.
      • et al.
      Effect of phenylephrine and ephedrine bolus treatment on cerebral oxygenation in anaesthetized patients.
      • Nissen P.
      • Brassard P.
      • Jørgensen T.B.
      • Secher N.H.
      Phenylephrine but not ephedrine reduces frontal lobe oxygenation following anesthesia-induced hypotension.
      One study that compared prophylactic phenylephrine versus prophylactic ephedrine during CD found that ephedrine administration was associated with maintenance of mean frontal lobe ScO2 compared to baseline (+2.1 ± 2.8%; mean ± SE), whereas phenylephrine was associated with decreased mean frontal lobe ScO2 (−8.6 ± 2.8%; P=0.005), with a 10.7% difference in mean frontal lobe ScO2 between groups (P=0.0106).
      • Foss V.T.
      • Christensen R.
      • Rokamp K.Z.
      • Nissen P.
      • Secher N.H.
      • Nielsen
      HB. Effect of phenylephrine vs. ephedrine on frontal lobe oxygenation during caesarean section with spinal anesthesia: an open label randomized controlled trial.
      The mechanism for decreased mean frontal lobe ScO2 produced by phenylephrine may be via decreased cardiac output or cerebral vasoconstriction. Phenylephrine is known to cause a dose-dependent decrease in cardiac output.
      • Orbach-Zinger S.
      • Bizman I.
      • Firman S.
      • et al.
      Perioperative noninvasive cardiac output monitoring in parturients undergoing cesarean delivery with spinal anesthesia and prophylactic phenylephrine drip: a prospective observational cohort study.
      In a previous study performed in our institution, measuring cardiac output at different time points during CD performed under spinal anesthesia with a prophylactic phenylephrine infusion, the pattern of changes in stroke volume at different time points corresponded to the changes in mean frontal lobe ScO2 seen in the current study.
      • Orbach-Zinger S.
      • Bizman I.
      • Firman S.
      • et al.
      Perioperative noninvasive cardiac output monitoring in parturients undergoing cesarean delivery with spinal anesthesia and prophylactic phenylephrine drip: a prospective observational cohort study.
      In the study comparing ephedrine and phenylephrine boluses, phenylephrine decreased cardiac output while ephedrine maintained it; and there was a significant correlation between mean frontal lobe ScO2 and cardiac output (P <0.0010).
      • Meng L.
      • Cannesson M.
      • Alexander B.S.
      • et al.
      Effect of phenylephrine and ephedrine bolus treatment on cerebral oxygenation in anaesthetized patients.
      Decreased cardiac output to the brain may cause a compensatory increase in cerebral oxygen extraction, leading to decreased mean frontal lobe ScO2.
      • Denault A.
      • Deschamps A.
      • Murkin J.M.
      A proposed algorithm for the intraoperative use of cerebral near-infrared spectroscopy.
      However, phenylephrine may cause a decrease in mean frontal lobe ScO2 via mechanisms other than those that change cardiac output. In a study of eight healthy volunteers receiving a phenylephrine infusion at one intervention and normal saline infusion at a different intervention, cardiac output, mean frontal lobe ScO2, middle cerebral artery mean blood velocity, and right internal carotid artery and internal jugular venous blood flow assessed using duplex ultrasound were measured. Compared to saline administration, phenylephrine infusion was associated with a decreased mean frontal lobe ScO2, an increased mean cerebral artery velocity, an increased internal jugular venous blood flow, and no changes in either cardiac output or internal carotid artery blood flow.
      • Ogoh S.
      • Sato K.
      • Fisher J.P.
      • Seifert T.
      • Overgaard M.
      • Secher N.H.
      The effect of phenylephrine on arterial and venous cerebral blood flow in healthy subjects.
      The authors postulated that these changes could be due to cerebral vasoconstriction.
      We noted that the mean frontal lobe ScO2 remained low even one hour after surgery. There are several possible explanations for this phenomenon. In a previous study in our institution measuring perioperative cardiac output changes after spinal anesthesia with a phenylephrine infusion, we found that both cardiac output and non-invasive blood pressure remained below baseline when measured in the PACU one hour after surgery.
      • Ogoh S.
      • Sato K.
      • Fisher J.P.
      • Seifert T.
      • Overgaard M.
      • Secher N.H.
      The effect of phenylephrine on arterial and venous cerebral blood flow in healthy subjects.
      Clinical experience in our hospital has shown that spinal block takes approximately two hours to resolve, and so some deafferentation which remained may have influenced NIRS readings.
      Our study has several limitations. Firstly, we did not measure cardiac output. We did not study short- and long-term neurological sequelae using cognitive tests in our patients. There were a few cases where patients with a low mean frontal lobe ScO2 complained of weakness, tiredness or agitation, but these complaints were not quantified. We did not measure time intervals between spinal initiation and delivery of the baby or progression of sensory levels intraoperatively or postoperatively. We did not give oxygen supplementation to women intraoperatively: in a previous study, oxygen supplementation was shown to be effective in attenuating the decrease in mean frontal lobe ScO2 readings after spinal anesthesia.
      • Hirose N.
      • Kondo Y.
      • Maeda T.
      • Suzuki T.
      • Yoshino A.
      • Katayama Y.
      Oxygen supplementation is effective in attenuating maternal cerebral blood deoxygenation after spinal anesthesia for cesarean section.
      Therefore our results may not be able to be extrapolated in centers that routinely give oxygen supplementation. Likewise, our fluid supplementation is different from others. Large variations in fluid supplementation exist,
      • Loubert C.
      Fluid and vasopressor management for Cesarean delivery under spinal anesthesia: continuing professional development.
      • Ngan Kee W.D.
      • Khaw K.S.
      • Ng F.F.
      Prevention of hypotension during spinal anesthesia for cesarean delivery: an effective technique using combination phenylephrine infusion and crystalloid cohydration.
      and the effect of crystalloid co-loading on hemodynamics is still debated.
      • Ngan Kee W.D.
      • Khaw K.S.
      • Ng F.F.
      Prevention of hypotension during spinal anesthesia for cesarean delivery: an effective technique using combination phenylephrine infusion and crystalloid cohydration.
      • McDonald S.
      • Fernando R.
      • Ashpole K.
      • Columb M.
      Maternal cardiac output changes after crystalloid or colloid coload following spinal anesthesia for elective cesarean delivery: a randomized controlled trial.
      • Mercier F.J.
      Cesarean delivery fluid management.
      • Banerjee A.
      • Stocche R.M.
      • Angle P.
      • Halpern S.H.
      Preload or coload for spinal anesthesia for elective Cesarean delivery: a meta-analysis.
      Finally, the mean frontal lobe ScO2 readings may be inaccurate, as a recent study showed NIRS readings were prone to extracranial contamination and a recent study has put into question the accuracy of NIRS measurement.
      • Greenberg S.
      • Murphy G.
      • Shear T.
      • et al.
      Extracranial contamination in the INVOS 5100C versus the FORE-SIGHT ELITE cerebral oximeter: a prospective observational crossover study in volunteers.
      • Kahn R.A.
      • Anyanwu A.
      Near-infrared spectroscopy in vegetables and humans: An observational study.
      In conclusion, we observed a fall in mean frontal lobe ScO2 in parturients undergoing CD under spinal anesthesia with a prophylactic phenylephrine infusion. This was observed at almost all points during surgery and at one hour postpartum, with one third of parturients having at least one mean frontal lobe ScO2 reading below the theoretical cerebral hypoperfusion level. Though the findings of this pilot study are interesting, it is not possible to generate conclusions. Larger randomized controlled studies are needed to elucidate the cause, and long-term outcome studies to assess its clinical significance.

      Declaration of Interests

      Dr Weiniger is an editor of the International Journal of Obstetric Anesthesia but took no role in the assessment of this submission. There was no funding support received for this research.

      Appendix A. Supplementary data

      The following are the Supplementary data to this article:

      References

        • Mercier F.J.
        • Auge M.
        • Hoffmann C.
        • Fischer C.
        • Le Gouez A.
        Maternal hypotension during spinal anesthesia for caesarean delivery.
        Minerva Anestesiol. 2013; 79: 62-73
        • Habib A.S.
        A review of the impact of phenylephrine administration on maternal hemodynamics and maternal and neonatal outcomes in women undergoing cesarean delivery under spinal anesthesia.
        Anesth Analg. 2012; 114: 377-390
        • Calderon-Arnulphi M.
        • Alaraj A.
        • Slavin K.V.
        Near infrared technology in neuroscience: past, present and future.
        Neurol Res. 2009; 31: 605-614
        • Sun S.
        • Liu N.H.
        • Huang S.Q.
        Role of cerebral oxygenation for prediction of hypotension after spinal anesthesia for caesarean section.
        J Clin Monit Comput. 2016; 30: 417-421
        • Hirose N.
        • Kondo Y.
        • Maeda T.
        • Suzuki T.
        • Yoshino A.
        Relationship between regional cerebral blood volume and oxygenation and blood pressure during spinal anesthesia in women undergoing cesarean section.
        J Anesth. 2016; 30: 603-609
        • Berlac P.A.
        • Rasmussen Y.H.
        Per-operative cerebral near-infrared spectroscopy (NIRS) predicts maternal hypotension during elective caesarean delivery in spinal anaesthesia.
        Int J Obstet Anesth. 2005; 14: 26-31
        • Foss V.T.
        • Christensen R.
        • Rokamp K.Z.
        • Nissen P.
        • Secher N.H.
        • Nielsen
        HB. Effect of phenylephrine vs. ephedrine on frontal lobe oxygenation during caesarean section with spinal anesthesia: an open label randomized controlled trial.
        Front Physiol. 2014; 5: 81
        • Loubert C.
        Fluid and vasopressor management for Cesarean delivery under spinal anesthesia: continuing professional development.
        Can J Anaesth. 2012; 59: 604-619
        • Kinsella S.M.
        • Carvalho B.
        • Dyer R.A.
        • et al.
        International consensus statement on the management of hypotension with vasopressors during caesarean section under spinal anaesthesia.
        Anaesthesia. 2018; 73: 71-92
        • Meng L.
        • Cannesson M.
        • Alexander B.S.
        • et al.
        Effect of phenylephrine and ephedrine bolus treatment on cerebral oxygenation in anaesthetized patients.
        Br J Anaesth. 2011; 107: 209-217
        • Nissen P.
        • Brassard P.
        • Jørgensen T.B.
        • Secher N.H.
        Phenylephrine but not ephedrine reduces frontal lobe oxygenation following anesthesia-induced hypotension.
        Neurocrit Care. 2010; 12: 17-23
        • Clemmesen C.G.
        • Pedersen L.M.
        • Hougaard S.
        • et al.
        Cerebral oximetry during preoperative resuscitation in elderly patients with hip fracture: a prospective observational study.
        J Clin Monit Comput. 2018; https://doi.org/10.1007/s10877-018-0107-6
        • Casati A.
        • Fanelli G.
        • Pietropaoli P.
        • et al.
        Continuous monitoring of cerebral oxygen saturation in elderly patients undergoing major abdominal surgery minimizes brain exposure to potential hypoxia.
        Anesth Analg. 2005; 101: 740-747
        • Murkin J.M.
        • Adams S.J.
        • Novick R.J.
        • et al.
        Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study.
        Anesth Analg. 2007; 104: 51-58
        • Fassoulaki A.
        • Paraskeva A.
        • Tsaroucha A.
        Cesarean delivery under spinal anesthesia is associated with decreases in cerebral oxygen saturation as assessed by NIRS: an observational study.
        Cur Med Res Opin. 2014; 30: 331-337
        • Ilies C.
        • Kiskalt H.
        • Siedenhans D.
        • et al.
        Detection of hypotension during Caesarean section with continuous non-invasive arterial pressure device or intermittent oscillometric arterial pressure measurement.
        Br J Anaesth. 2012; 109: 413-419
        • Sadato N.
        • Zeffiro T.A.
        • Campbell G.
        • Konishi J.
        • Shibasaki H.
        • Hallett M.
        Regional cerebral blood flow changes in motor cortical areas after transient anesthesia of the forearm.
        Ann Neurol. 1995; 37: 74-81
        • Ogoh S.
        • Sato K.
        • Fisher J.P.
        • Seifert T.
        • Overgaard M.
        • Secher N.H.
        The effect of phenylephrine on arterial and venous cerebral blood flow in healthy subjects.
        Clin Physiol Funct Imaging. 2011; 31: 445-451
        • Orbach-Zinger S.
        • Bizman I.
        • Firman S.
        • et al.
        Perioperative noninvasive cardiac output monitoring in parturients undergoing cesarean delivery with spinal anesthesia and prophylactic phenylephrine drip: a prospective observational cohort study.
        J Matern Fetal Neonatal Med. 2018; 22: 1-7
        • Denault A.
        • Deschamps A.
        • Murkin J.M.
        A proposed algorithm for the intraoperative use of cerebral near-infrared spectroscopy.
        Semin Cardiothorac Vasc Anesth. 2007; 11: 274-281
        • Hirose N.
        • Kondo Y.
        • Maeda T.
        • Suzuki T.
        • Yoshino A.
        • Katayama Y.
        Oxygen supplementation is effective in attenuating maternal cerebral blood deoxygenation after spinal anesthesia for cesarean section.
        Adv Exp Med Biol. 2016; 876: 471-477
        • Ngan Kee W.D.
        • Khaw K.S.
        • Ng F.F.
        Prevention of hypotension during spinal anesthesia for cesarean delivery: an effective technique using combination phenylephrine infusion and crystalloid cohydration.
        Anesthesiology. 2005; 103: 744-750
        • McDonald S.
        • Fernando R.
        • Ashpole K.
        • Columb M.
        Maternal cardiac output changes after crystalloid or colloid coload following spinal anesthesia for elective cesarean delivery: a randomized controlled trial.
        Anesth Analg. 2011; 113: 803-810
        • Mercier F.J.
        Cesarean delivery fluid management.
        Curr Opin Anaesthesiol. 2012; 25: 286-291
        • Banerjee A.
        • Stocche R.M.
        • Angle P.
        • Halpern S.H.
        Preload or coload for spinal anesthesia for elective Cesarean delivery: a meta-analysis.
        Can J Anaesth. 2010; 57: 24-31
        • Greenberg S.
        • Murphy G.
        • Shear T.
        • et al.
        Extracranial contamination in the INVOS 5100C versus the FORE-SIGHT ELITE cerebral oximeter: a prospective observational crossover study in volunteers.
        Can J Anaesth. 2016; 63: 24-30
        • Kahn R.A.
        • Anyanwu A.
        Near-infrared spectroscopy in vegetables and humans: An observational study.
        Eur J Anaesthesiol. 2018; 35: 907-910https://doi.org/10.1097/EJA.00000000000855